An Enhanced Multi-Frequency Distorted Born Iterative Method for Ultrasound Tomography Based on Fundamental Tone and Overtones
Conventional Distorted Born Iterative Method (DBIM) using single frequency has low resolution and is prone to creating images with high-contrast subjects. We propose a productive frequency combination method to better result in tomographic ultrasound imaging based on the multi-frequency technique. T...
Lưu vào:
Tác giả chính: | , , , |
---|---|
Định dạng: | Bài trích |
Ngôn ngữ: | English |
Nhà xuất bản: |
IGI Global
2022
|
Chủ đề: | |
Truy cập trực tuyến: | https://www.igi-global.com/gateway/article/289608 https://dlib.phenikaa-uni.edu.vn/handle/PNK/5849 https://doi.org/10.4018/ijirr.289608 |
Từ khóa: |
Thêm từ khóa
Không có từ khóa, Hãy là người đầu tiên đánh dấu biểu ghi này!
|
Tóm tắt: | Conventional Distorted Born Iterative Method (DBIM) using single frequency has low resolution and is prone to creating images with high-contrast subjects. We propose a productive frequency combination method to better result in tomographic ultrasound imaging based on the multi-frequency technique. This study uses the natural mechanism of emitting oscillators' frequencies and uses these frequencies for imaging in iterations. We use a fundamental tone (i.e., the starting frequency f0) for the first iteration in DBIM, then consecutively use its overtones for the next ones. The digital simulation scenarios are tested with other multi-frequency approaches to prove our method's feasibility. We performed 57 different simulation scenarios on the use of multi-frequency information for the DBIM method. As a result, the proposed method for the smallest normalization error (RRE = 0.757). The proposed method's imaging time is not significantly longer than the way of using single frequency information. |
---|